FLIR CM174 USER MANUAL Document Identifier: CM174-en-US_AC
32
7.2 IR Thermal Imaging Specifications
Detector type FLIR LeptonTM; Micro-bolometer Focal Plane Array (FPA)
Thermal Sensitivity 150mK
IR Imaging resolution 60 x 80 pixels
IR Imaging field of view 50.0 o x 38.6 o (vertical x horizontal)
IR Imaging spectral response 8 to 14μm
IR Image Capture Frequency 9Hz
IR Image Color Palettes Programmable: Iron, Rainbow, and Gray scale
Shutter Integrated, automatic shutter
Laser Pointer Class I (red)
IR Temperature Measurement range 14°F ~ +302°F (-10°C ~ +150°C)
Over- and under- range indication OL
Temperature reading stabilization Dashes are displayed for approximately 30 seconds as
temperature reading stabilizes
IR Temperature Resolution 0.1°F (0.1°C)
IR Temperature Accuracy 5.4°F (±3°C) or ± 3% of reading; whichever is greater (after
2-minute warm-up period)
Distance to Spot (D:S) ratio 30:1
Temperature Scanning Continuous
Emissivity 4 presets plus a custom setting (0.10 to 0.99)
Targeting Displayed Cross hairs pinpoint center of measurement spot