5.10.2 Programming Menu 
1.  To access the programming menu, set the function switch to any position to switch 
the meter on, and then press and hold the IGM button for > 3 seconds. The menu 
should now show a display, similar to the one in Fig. 5-11. 
 Figure 5.11 Programming menu example 
2.  The menu icons are located on the left side of the display and their settings are 
shown on the right side. Use the up and down arrow buttons to scroll the parameter 
list. 
3.  When a parameter is highlighted, press the OK button to either toggle the setting (if 
it is a parameter with two options: Laser ON/OFF, Temperature units C/F, Cross hairs 
ON/OFF, Diode mode SMART/CLASSIC) or to open the parameter for editing (for 
parameters with 3 or more selection options: Color Palette, Emissivity, APO). For 
details on each, refer to the following paragraphs. 
4.  Once programming is complete, press the IGM button momentarily to exit the 
thermal imaging mode. 
Figure 5.12 Programming menu ICON descriptions 
Presets: 0.30, 0.60, 0.80, 0.95 (and custom) 
OFF, 1, 2, 5, or 10 minutes