35
Table 7.4 Inrush current
Function
Range
Accuracy
ACA Inrush
60.00 A
±(3% + 0.3 A)
600.0 A
±(3% + 5 dgt)
Inrush current detection threshold: 0.5A for 60A range and 5.0A for 600.0A range
Overload protection: 1000 V (rms), 600 A (rms).
Integration time is 100 ms.
Table 7.5 Resistance and continuity and diode
Function
Range
Accuracy
Resistance
600.0 Ω
±(1.0% + 5 dgt)
6.000 KΩ
Continuity
600.0 Ω
±(1.0% + 5 dgt)
Diode
1.5V
±(1.5% + 5 dgt)
Overload protection: 1000 V (rms).
Maximum test current: Approx. 0.1 mA.
Maximum open circuit voltage for Ω: Approx. 1.8 V.
Maximum open circuit voltage for diode: Approx. 1.8 V.
Continuity threshold:
< 30 Ω beeper on.
> 150 Ω beeper off.
Continuity indicator: 2.7 kHz tone buzzer.
Continuity response time: <100 ms.