True RMS Digital Multimeters
Operating Basics
21
This calibrates the meter for the operating environment.
t
shows in the upper
display. If you hear an error beep, the offset is greater than ±5 °C. You can
repeat this calibration at any time. To undo this calibration, return the meter to
factory settings by pressing both the yellow (shift) button and
MIN MAX while
powering up the meter (see Table 3. Power-up Options).
To insure accuracy of temperature measurements, you should follow this
procedure when using other K-type thermocouple probes with the 183/185
DMMs because accuracy specifications vary in different types of probes.
Note
Observe proper polarity on the probe adapter and do not calibrate
the offset immediately following high amperage measurements.
Current Measurements
Measurement Softkey Connect leads
Main
Display
Upper
Display
True RMS AC
Amps
1 AC AC Hz
DC Amps
(default)
2 DC DC --
Amps AC DC
dual display
DC AC
Amps AC+DC
total RMS
1
3 AC
DC or
AC+DC
(press to
toggle)
AC+DC Hz
4-20 mA
current %
2
(185 only)
4
20 mA
LoadCurrent
source
ahr17f.eps
DC %
1
22
IDCIACIrms +=
2
4-20 mA measurement is used in process loop calibration. % = (measured
current – 4 mA)/16 mA