183 & 185
Users Manual
32
Once you set the meter to MIN MAX mode, the meter will uprange, or
automatically increase range setting, but will not downrange. You must exit
MIN MAX mode to restore downranging abilities to the meter.
While in MIN MAX mode, press
HOLD to stop the recording and freeze the
most recent MIN MAX values. Press
MIN MAX to cycle through the held
values. Press
HOLD a second time to reset and restart the MIN MAX
recording.
While in MIN MAX mode, activate memory mode to hold the displayed MIN
MAX value for storage into memory without stopping or resetting the MIN
MAX recording.
To exit MIN MAX mode and clear stored values, press and hold
MIN MAX for
two seconds.
Auto Fuse Detection
The meter automatically verifies the integrity of the internal fuse when you set
the measurement function knob to A. If an open fuse is detected, the word
FUSE
shows on the main display.
See "Replacing the Fuse" for instructions on how to replace fuse F1.