True RMS Digital Multimeters
Specifications
39
Table 14. Resistance Range, Resolution, and Accuracy
Resolution Accuracy
Range 5,000 Counts 50,000 Counts 183 185
50 0.01
1
–– ± (0.1% + 10
counts)
500 0.1 0.01 Ω± (0.1% + 4
counts)
2
5 k 1 0.1 Ω± (0.1% + 2
counts)
3
50 k 10 1
500 k 100 10
5 M 1 k 100 Ω± (0.4% + 4
counts)
2
50 M 10 k –– ± (1.0% + 4
counts)
2
Temperature coefficient 50 to 500 k: Add (0.03% + 0.1 ct.)/°C
beyond rated temperature range.
5 M to 50 M: Add (0.2% + 0.1 ct.)/°C
beyond rated temperature range.
1
5,000 count mode only.
2
Accuracy in 50,000-count mode is % + 40 counts.
3
Accuracy in 50,000-count mode is % + 20 counts.
Table 15. Continuity Characteristics
Characteristics Description
Continuity threshold
Beeper sounds when resistance is 100 or
less (typical)
Response time < 1 ms