183 & 185
Users Manual
34
Table 5. DC Voltage Characteristics
Characteristic Description
Settling time 3 readings (typical)
Reading rate 5,000 ct.: 4 readings per second
50,000 ct.: 1 reading per second
Rejection ratio
Common mode 120 dB at DC or 50 Hz or 60 Hz
Normal mode 60 dB at 50 Hz or 60 Hz
Input impedance 10 M (typical)
Table 6. DC Voltage Range, Resolution, and Accuracy
Resolution Accuracy
1
Range
5,000
counts
50,000
counts 183 185
0.5 V 100 µV 10 µV
5 V 1 mV 100 µV
50 V 10 mV 1 mV
500 V 100 mV 10 mV
±(0.07% + 1
count)
± (0.05% + 1
count)
1000 V 1 V 100 mV
Temperature coefficient Add (0.005% + 0.1 ct.)/°C to accuracy
beyond rated temperature range.
1
Accuracy in 50,000-count mode is % + 10 counts.