Fluke 183 - Frequency Characteristics, Resolution, and Accuracy; Frequency Voltage Range; Duty Factor Characteristics

Fluke 183
60 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
True RMS Digital Multimeters
Specifications
41
Table 18. Frequency Characteristics, Resolution, and Accuracy
Characteristics Description
Signal coupling AC
Minimum frequency 0.5 Hz
Maximum frequency 1 MHz
Accuracy ± (0.002%) + 1 count
Best resolution 10,000 count: 0.01 Hz
100,000 count: 0.001 Hz
Temperature coefficient Add 0.00004%/°C beyond rated temperature
range.
Table 19. Frequency Voltage Range
Range Sensitivity, 10 Hz - 100 kHz Sensitivity, 1 MHz
1
500 mV 100 mV ––
5 V 500 mV 2 V
50 V 5 V 20 V
500 V 50 V ––
1
For voltages > 100 V, the maximum volts–Hz the meter is < 10
7
VHz.
Table 20. Duty Factor Characteristics
Characteristics Description
Range 1 Hz to 100 kHz
Accuracy ± (0.1% + 0.05% per kHz) for 5 V input (logic
signals only)
Signal coupling DC
Resolution 0.1%
Sensitivity 30% of range