True RMS Digital Multimeters
Specifications
37
Table 10. DC Current Range, Resolution, and Accuracy
Resolution Accuracy
Range 5,000 Counts
50,000
Counts 183 185
500 µA 100 nA 10 nA ± (0.2% + 4 counts)
1
5 mA 1 µA 100 nA
50 mA 10 µA1 µA
500 mA 100 µA 10 µA
± (0.2% + 2 counts)
2
5 A 1 mA 100 µA ± (0.4% + 2 counts)
2
10 A for 3 minutes
(15 A for 30 sec.)
Allow 10 minutes of
cooling between
measurements
10 mA 1 mA ± (0.8% + 2 counts)
2
Temperature
coefficient
Add (0.05% + 0.1 ct.)/°C beyond rated
temperature range.
1
Accuracy in 50,000-count mode is % + 40 counts.
2
Accuracy in 50,000-count mode is % + 20 counts.
Table 11. AC Current Characteristics
Characteristics Description
Burden voltage 0.5 mA to 5 A: 0.9 V max.
10 A: 1.0 V max.
AC+DC
1
Accuracy AC RMS amps accuracy + DC amps
accuracy
Upper display frequency
Accuracy ± (0.002% + 1 count) for 20 Hz to 5 kHz
Sensitivity 10% of range
Settling time 4 readings (typical)
Reading rate 5,000 ct.: 4 readings per second
50,000 ct.: 1 reading per second
1
5,000-count mode only.