EasyManua.ls Logo

Fluke 183 - Temperature Characteristics; FAST MIN MAX (1 Ms) Characteristics; Physical Characteristics

Fluke 183
60 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
183 & 185
Users Manual
42
Table 21. Temperature Characteristics
Characteristics Description
Main display
Range –50 °C to +980 °C
Accuracy ± 3 °C
1
(typical)
Thermocouple type K
Upper display
Accuracy ± 3 °C of ambient temperature (typical)
1
Use the water and ice offset calibration method in the "Temperature
Measurements (185 only)" section for accuracy to ±1.0 °C.
Table 22. FAST MIN MAX (1 ms)Characteristics
Characteristics Description
Accuracy
1
Specified voltage or current measurement
± 30 counts of the peak value of a single 1ms
pulse.
1
5,000-count mode only.
Table 23. Physical Characteristics
Characteristic Description
Dimensions (H × W × D) 38 mm × 88 mm × 183 mm (without holster)
Weight (with batteries) 383 g (13.5 oz)
With holster 539 g (1 lb 3 oz)