Fluke 190-104/190-204
Service Information
4-10
8. Next check channel B, C or D in succession:
Press
TRIGGER
and select B, C or D as trigger source with
F1
,
Press B, C or D to assign vertical ‘range’ and ‘move’ to channel B, C or D,
Observe reading B, C or D.
9. When you are finished, set the 5500A to 0 (zero) Volt, and to Standby.
Table 4-2. Volts DC Measurement Verification Points
Range 5500A output V dc Input A, B, C, D Reading
2 mV/div +6.0 mV +4.9 to +7.1
-6.0 mV -4.9 to -7.1
5 mV/div +15.0 mV +14.3 to +15.7
-15.0 mV -14.3 to -15.7
10 mV/div +30.0 mV +29.1 to +30.9
-30.0 mV -29.1 to -30.9
20 mV/div +60.0 mV +58.6 to +61.4
-60.0 mV -58.6 to -61.4
50 mV/div +150 mV +143 to +157
-150 mV -143 to -157
100 mV/div +300 mV +291 to +309
-300 mV -291 to -309
200 mV/div +600 mV +586 to +614
-600 mV -586 to -614
500 mV/div +1.50 V +1.43 to +1.57
-1.50 V -1.43 to -1.57
1 V/div +3.00 V +2.91 to +3.09
-3.00 V -2.91 to -3.09
2 V/div +6.00 V +5.86 to +6.14
-6.00 V -5.86 to -6.14
5 V/div +15.0 V +14.3 to +15.7
-15.0 V -14.3 to -15.7
10 V/div +30.0 V +29.1 to +30.9
-30.0 V -29.1 to -30.9
20 V/div +60.0 V +58.6 to +61.4
-60.0 V -58.6 to -61.4
50 V/div +150 V +143 to +157
-150 V -143 to -157