Calibration Adjustment
5.6 Final Calibration 5
5-9
ADC
B
NORMAL
SCOPE
FLUKE 5500A CALIBRATOR
PM9091
PM9585 - 50 OHM
USE 50 OHM
TERMINATION
Figure 5-3. 5500A SCOPE Output to Test Tool Input A.
5.6.2 Input A LF-HF Gain
Proceed as follows to do the Input A LF-HF Gain calibration:
1. Connect the Test Tool to the 5500A as shown in Figure 5-3.
2. The display must show step CL 0654. If it does not, then press
F1
or
F2
to select the first calibration step in Table 5-1.
3. Set the 5500A SCOPE output to source the signal required for the first calibration
point in Table 5-1.
4. Set the 5500A in operate (OPR) or standby (STBY) as indicated.
5. Press
F3
to start the calibration.
6. Wait until the display shows calibration status
:READY .
7. Press
F2
to select the next calibration step, set the 5500A to the next
calibration point signal, and start the calibration.
Continue through all calibration points of Table 5-1.
8. When you are finished, set the 5500A to Standby.
9. Continue at Section 5.6.3.