EasyManua.ls Logo

Fluke 190-104 - Input D LF-HF Gain

Fluke 190-104
80 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Fluke 190-104/190-204
Service Information
5-12
Table 5-3. Input C LF-HF Gain Calibration Points
Cal step UUT input signal 5500A Setting
CL 0656 0.5 Vpp square wave, 1 kHz SCOPE edge, 0.5 Vpp, 1 kHz
CL 0470 0.5 Vpp square wave, 1 kHz SCOPE edge, 0.5 Vpp, 1 kHz
CL 0714 0.5 Vpp square wave, 1 kHz SCOPE edge, 0.5 Vpp, 1 kHz
CL 0478 0.5 Vpp square wave, 1 kHz SCOPE edge, 0.5 Vpp, 1 kHz
CL 0484 0.62 Vpp sine wave, 50 kHz SCOPE levsine, 0.62 Vpp, 50 kHz
CL 0485 0.62 Vpp sine wave
Fluke 190-204: 221 MHz
Fluke 190-104: 141 MHz
SCOPE levsine, 0.62 Vpp,
221 MHz
141 MHz
ADC
B
NORMAL
SCOPE
FLUKE 5500A CALIBRATOR
PM9091
PM9585 - 50 OHM
USE 50 OHM
TERMINATION
Figure 5-5. 5500A SCOPE Output to Test Tool Input C
5.6.5 Input D LF-HF Gain
Proceed as follows to do the Input D LF-HF Gain calibration:
1. Press
F2
to select the first calibration step in Table 5-4.
2. Connect the Test Tool to the 5500A as shown in Figure 5-6.
3. Set the 5500A SCOPE output to source the signal required for the first calibration
point in Table 5-4.
4. Set the 5500A in operate (OPR) or standby (STBY) as indicated.
5. Press
F3
to start the calibration.
6. Wait until the display shows calibration status
:READY .
7. Press
F2
to select the next calibration step, set the 5500A to the next
calibration point signal, and start the calibration.
Continue through all calibration points of Table 5-4.

Table of Contents

Related product manuals