EasyManua.ls Logo

Fluke 190-104 - Input D Trigger Sensitivity Test

Fluke 190-104
80 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Fluke 190-104/190-204
Service Information
4-26
4.6.15 Input D Trigger Sensitivity Test
Proceed as follows to test the Input D trigger sensitivity:
1. Connect the Test Tool to the 5500A as shown in Figure 4-10.
ADC
B
NORMAL
SCOPE
FLUKE 5500A CALIBRATOR
PM9091
PM9585 - 50 OHM
USE 50 OHM
TERMINATION
Figure 4-10. 5500A Scope Output to Test Tool Input D
2. Select the following Test Tool setup:
Press
D
and use
F1
to turn Input D on.
Press
C
and use
F1
to turn Input C off.
Press
D
: using
MOVE
move the Input D trace zero to the center grid line.
Press
TRIGGER
and use
F1
to select Input D as trigger source.
Using
mV
RANGE
and
V
RANGE
change the sensitivity range to select manual sensitivity
ranging, and lock the Input D sensitivity range on 2 V/div.
3. Using
TIMEsns
select the time base indicated under the first column of Table 4-14.
4. Set the 5500A to source the leveled sine wave given in the first row of Table 4-14.
5. Adjust the 5500A output voltage until the displayed trace has the amplitude indicated
under the appropriate column of Table 4-14.
6. Verify that the signal is well triggered.
If it is not, press
TRIGGER
, then using
F3
enable the up/down arrow keys for
manual Trigger Level adjustment. Adjust the trigger level and verify that the signal
will be triggered now. The trigger level is indicated by the trigger icon (
).
7. Continue through the test points.
8. When you are finished, set the 5500A to Standby.
Table 4-14. Input B Trigger Sensitivity Test Points
UUT UUT 5500A SC... MODE levsin UUT
Model Time base Initial Input Voltage Frequency Trigger Amplitude
ALL 200 ns/div 100 mV pp 5 MHz 0.5 div

Table of Contents

Related product manuals