Calibration Adjustment
5.7 Save Calibration Data and Exit 5
5-17
Table 5-6. Input ABCD Gain Calibration Points
Cal step UUT input value (5500A NORMAL)
CL 0799 5 mV
CL 0800 12.5 mV
CL 0801 25 mV
CL 0802 50 mV
CL 0803 125 mV
CL 804 250 mV
CL 0805 500 mV
CL 0806 1.25 V
CL 0807 2.5 V
CL 0808 5 V
CL 0809 12.5 V
CL 0810 25 V
CL 0811 50 V (set 5500A to OPR!)
CL 0812 125 V
CL 0813 250 V
5.6.9 Input ABCD Zero
Proceed as follows to do the Input ABCD Zero calibration:
1. Press
F2
to select calibration adjustment step CL0852
2. Short circuit Input A, B, C, D.
3. Press
F3
to start the zero calibration
4. Wait until the display shows the status
:READY.
5. Remove the input terminations.
6. Continue at Section 5.7.
5.7 Save Calibration Data and Exit
Proceed as follows to save the calibration data, and to exit the Maintenance mode:
1. Remove all test leads from the Test Tool inputs.
2. Press
F4
EXIT. The Test Tool will display: