71X Series
Calibration Manual
4
To avoid overpressure damage, do not apply pressure
that exceeds limits listed in the Users Manual for the
specific product.
713, 717 and 718 only: To avoid corrosion in the
pressure sensor, use the calibrator only with media
compatible with glass, ceramic, silicon, RTV, nitrile,
(Buna -N) type 303 stainless steel, and nickel.
718 and 718Ex only: To avoid damage to the pump, use
with dry air and non-corrosive gases only. Use of the
optional Fluke 700-ILF In-Line Filter may help isolate the
pump from contaminates.
Hold in
fixed
position
kf01f.eps
Figure 1. Proper Use of Tools (713, 717, 718, and 718Ex Models)