Process Calibrators
Required Equipment
25
Snaps
F1
Wh003f.eps
Figure 6. Replacing the Fuse (718 shown)
Required Equipment
The equipment required to perform the verification and calibration procedures in this
manual is identified in Table 34.
Table 34. Required Calibration Equipment
Calibrator
Model
Equipment Minimum Specifications
Recommended
Model
712 DC Calibrator 5 to 3000 Ω
Accuracy: 70 ppm + 0.025 Ω
Fluke 5520A
Reference Multimeter 5 to 3000 Ω
Accuracy: 70 ppm + 0.037 Ω
Fluke 8508A
Four Test Leads 5500A/LEADS
713 DC Calibrator 0 to 24 mA Fluke 5520A
Reference Multimeter 0 to 24 mA
Accuracy: 0.006 % + 0.25 mA
Fluke 8508A
Dead Weight Tester -12 to 100 PSIG
Accuracy: 0.012 % of range
Teflon Tape
Two Test Leads 5500A/LEADS