71X Series
Calibration Manual
32
Table 43. 713 Pressure Verification
713 30G 713 100G
Input pressure Display Reading Input pressure Display Reading
0 psi -0.015 to 0.015 0 psi -0.05 to 0.05
6 psi 5.985 to 6.015 20 psi 19.95 to 20.05
12 psi 11.985 tp 12.015 40 psi 39.95 to 40.05
18 psi 17.985 to 18.015 60 psi 59.95 to 60.05
24 psi 23.985 to 24.015 80 psi 79.95 to 80.05
30 psi 29.985 to 30.015 100 psi 99.95 to 100.05
24 psi 23.985 to 24.015 80 psi 79.95 to 80.05
18 psi 17.985 to 18.015 60 psi 59.95 to 60.05
12 psi 11.985 tp 12.015 40 psi 39.95 to 40.05
6 psi 5.985 to 6.015 20 psi 19.95 to 20.05
-12 psi * -12.015 to -11.985 -12 psi* -12.05 to -11.95
* Vacuum is only specified for V1.3 and later. To display the firmware version, start with the 713 off, press and
hold M, then press the power button.
mA Measure Verification
The display should indicate:
0.000mA 0.00 PSI
1. Connect a test lead from Red AUX terminal of the 5520A to the Fluke 8508A I+
terminal.
2. Connect a test lead from Black AUX terminal of the 5520A to the 713 Calibrator
Com terminal.
3. Connect a test lead from I- terminal of the Fluke 8508A to the 713 Calibrator mA
terminal.
4. Adjust the 5520A if necessary so that the current shown on the Fluke 8508A is the
same as the 5520A values shown in Table 44.
5. Verify that the display readings on the 713 calibrator are within the limits.
Table 44. 713 mA Measure Verification
Fluke 5520A Fluke 713
4.0000 mA 3.998 mA to 4.002 mA
12.0000 mA 11.996 mA to 12.004 mA
24.0000 mA 23.993 mA to 24.007 mA
6. Disconnect the test leads and turn off the 713 Calibrator.