EasyManua.ls Logo

Fluke 716 - Ma Measure Verification

Fluke 716
88 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
71X Series
Calibration Manual
46
717 1500G 717 3000G
Input
Pressure
Display
Reading
Input
Pressure
Display
Reading
0
p
si 0.8 to -0.8 0
p
si 1.5 to -1.5
300
p
si 300.8 to 299.3 600.0
p
si 601.5 to 598.5
600
p
si 600.8 to 599.3 1200.0
si 1201.5 to 1198.5
900
p
si 900.8 to 899.3 1800.0
si 1801.5 to 1798.5
1200
p
si 1200.8 to 1199.3 2400.0
si 2401.5 to 2398.5
1500
p
si 1500.8 to 1499.3 3000.0
si 3001.5 to 2998.5
1200
p
si 1200.8 to 1199.3 2400.0
p
si 2401.5 to 2398.5
900
p
si 900.8 to 899.3 1800.0
si 1801.5 to 1798.5
600
p
si 600.8 to 599.3 1200.0
si 1201.5 to 1198.5
300
p
si 300.8 to 299.3 600.0
p
si 601.5 to 598.5
717 5000G
Input
Pressure
Display
Reading
0 psi 2.5 to -2.5
1000 psi 1002.5 to 997.5
2000 psi 2002.5 to 1997.5
3000 psi 3002.5 to 2997.5
4000 psi 4002.5 to 3997.5
5000 psi 5002.5 to 4997.5
4000 psi 4002.5 to 3997.5
3000 psi 3002.5 to 2997.5
2000 psi 2002.5 to 1997.5
1000 psi 1002.5 to 997.5
*Vacuum is only specified for V1.3 and later. To display the firmware version, start with the 717 off, press
and hold M, then press the power button.
4. Carefully vent all pressure and disconnect the 717 from the deadweight tester.
mA Measure Verification
1. Press
O to turn power on. The display should read:
---------- mA
then change to:
0.000 mA 0.00PSI
2. Connect a test lead from Red AUX terminal of the 5520A to the Fluke 8508A I+
terminal.
3. Connect a test lead from Black AUX terminal of the 5520A to the 717 Calibrator
Com terminal.
4. Connect a test lead from I- terminal of the Fluke 8508A to the 717 Calibrator mA
terminal.

Table of Contents

Other manuals for Fluke 716

Related product manuals