Process Calibrators
Specifications
17
717 Specifications
Accuracy is specified for 1 year after calibration at operating temperatures of -10 °C to +
55 °C. To display firmware version, start with the unit off and press and hold M, then
press O.
Pressure
Table 25. 717 Pressure Specifications
Version 1.2 or Lower
Model Range SI Range Metric Max SI Max Metric
717-30 G (0 to 30) PSI 0 to 206.85 kPa 90 PSI 620 kPa
Version 1.3 or Higher
Model Range SI Range Metric Max SI Max Metric
717-1G (-1 to 1) PSI (-7 to 7) kPa 5 PSI 34.5 kPa
717-30G (-12 to 30) PSI (-83 to 207) kPa 60 PSI 413 kPa
717-100G (-12 to 100) PSI (-83 to 690) kPa 200 PSI 1379 kPa or 1.4 mPa
717-300G (-12 to 300) PSI
(-83 to 2068) kPa or
2.1 mPa 375 PSI 2586 kPa or 2.6 mPa
717-500G (0 to 500) PSI 3447 kPa or 3.4 mPa 1000 PSI 6895 kPa or 6.9 mPa
717-1000G (0-1000) PSI 6895 kPa or 6.9 mPa 2000 PSI 13790 kPa or 13.8 mPa
717-1500G (0-1500) PSI 10342 kPa or 10.3 mPa 3000 PSI 20684 kPa or 20.7 mPa
717-3000G (0-3000) PSI 20684 kpa or 20.7 mPa 6000 PSI 41369 kPa or 41.4 mPa
717-5000G (0-5000) PSI 34474 kPa or 34.5 mPa 10000 PSI 68948 kPa or 69 mPa
Accuracy: Pressure Accuracy is +/- 0.05% of range
Temperature coefficient: 0.01 % of range per °C for temperature ranges of between −10 °C to 18 °C and 28
°C to 55 °C.
Table 26. Pressure Display, Pressure Module Input
Range Resolution Accuracy
Refer to the Instruction Sheet for the pressure module
Table 27. DC mA Input
Range Resolution Accuracy,± ( % of Reading + Counts)
717 Version 1.2 24 mA 0.001 mA 0.025 + 1
717 Version 1.3 24 mA 0.001 mA 0.015 + 2