Process Calibrators
Calibration
63
V OUTPUT
1. In V OUTPUT mode, the unit displays “C 0” for 1 second followed by “ .0000”.
The Calibrator is now sourcing V using the default calibration.
2. Connect the test leads from V and COM to V and COM of ….DMM. The 715 should
source 0.000V +/- 0.010V.
3. Adjust the source value using the up and down big arrows for gross and small arrows
for fine adjustment, until you read 0.000V on the DMM.
4. Press
A to go to the next calibration point. “C 20” will be displayed for 1 second
and then “ .0000”. The 715 should source 20V with 5% tolerance.
5. Adjust the source value until you read 20.000 on the DMM.
6. Press
A to complete V OUTPUT calibration. “CHEC” will be displayed for 1
second. The new calibration constants are applied and the V OUTPUT calibration
can be checked.
7. Press
A to go to the next mode.
mV OUTPUT
1. In mV OUTPUT mode, the unit displays “C 0” for 1 second followed by “ 0.000”.
The calibrator is now sourcing mV using the default calibration. 715 should source
0.00mV +/- 0.10mV.
2. Adjust the source value using the up and down big arrows for gross and small arrows
for fine, until you read 0.00mV on the DMM.
3. Press
A to go to the next calibration point. “C 200” will be displayed for 1 second
and then “ 0.000”. The 715 should source 200.00mV with 5% tolerance.
4. Adjust the source value until you read 200.00 mV on the DMM.
5. Press
A to complete mV OUTPUT calibration. “CHEC” will be displayed for 1
second. The new calibration constants are applied and the mV OUTPUT calibration
can be checked.
6. Press
A to go to the next mode.
mA OUTPUT
1. In mA OUTPUT mode, the unit displays “C 1” for 1 second followe by “ .0000”.
The calibrator is now sourcing mA using the default calibration.
2. Connect the test leads from LOOP (red) and V (blak) to mA and COM jacks of
….DMM. The 715 should source 1.000mA +/- 0.010mA.
3. Adjust the source value using the up and down big arrows for gross and small arrows
for fine adjustment, until you read 1.000mA on the DMM.
4. Press
A to go to the next calibration point. “C 21” will be displayed for 1 second
and then “ .0000”. The 715 should source 21mA with 5% tolerance.
5. Adjust the source value until you read 21.000mA on the DMM.
6. Press
A to complete mA OUTPUT calibration. “CHEC” will be displayed for 1
second. The new calibration constants are applied and the mA OUTPUT calibration
can be checked.
7. Press
A to complete the calibration and go to the normal mode.

Table of Contents

Other manuals for Fluke 716

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 716 and is the answer not in the manual?

Fluke 716 Specifications

General IconGeneral
BrandFluke
Model716
CategoryTest Equipment
LanguageEnglish

Related product manuals