Process Calibrators
Specifications
13
Table 16. 714 Millivolt Measure and Source
Range Resolution Accuracy
-10 mV to 75 mV 0.01 mV ± (0.025 % of range (75 mV) + 1 count)
Maximum input voltage: 30 V
Firmware V2.0 and Later
Table 17. 714 Temperature Measure and Thermocouple Simulate
TC Type Range °C Accuracy °C *
-210.0 to 0.0 0.6
0.0 to 800.0 0.4
J
800.0 to 1200.0 0.5
-200.0 to 0.0 0.8
0.0 to 1000.0 0.5
K
1000.0 to 1372.0 0.7
-250.0 to 0.0 0.8
T
0.0 to 400.0 0.4
-250.0 to -100.0 0.8
E
-100.0 to 1000.0 0.4
-20.0 to 0.0 2.0
R
0.0 to 1787.0 1.4
-20.0 to 0.0 2.0
S
0.0 to 1767.0 1.4
600.0 to 800.0 1.4
800.0 to 1000.0 1.5
B
1000.0 to 1820.0 1.7
-200.0 to 0.0 0.45
L
0.0 to 900.0 0.4
-200.0 to 0.0 0.7
U
0.0 to 600.0 0.45
Maximum input voltage: 30 V
* Includes 0.2 °C cold junction compensation (CJC) error
Table 18. 714 Millivolt Measure and Source
Range Resolution Accuracy
-10 mV to 75 mV 0.001 mV 0.015 % ± 10 µV
Maximum input voltage: 30 V
Maximum source current is 1.0 mA