71X Series
Calibration Manual
18
Table 28. 717 Range and Resolution
1 30 100 300 500 1000 1500 3000 5000
Range -1 to 1 -12 to 30 -12 to
100
-12 to
300
0 to 500 0 to
1000
0 to
1500
0 to
3000
0 to
5000
psi 1.0000 30.000 100.00 300.00 500.00 1000.0 1500.0 3000.0 5000.0
bar 0.0689 2.0684 6.8947 20.684 34.474 68.947 103.42 206.84 344.74
mbar 68.948 2068.4 6894.8 20684 34474 68948 NA NA NA
kPa 6.8948 206.84 689.48 2068.4 3447.4 6894.8 10342 20684 34474
kg/cm2 0.0703 2.1092 7.0307 21.092 35.153 70.307 105.46 210.92 351.53
cmH2O@
4 °C
70.309 2109.3 7030.9 21093 35154 70309 NA NA NA
cmH2O@
20 °C
70.434 2113.0 7043.4 21130 35217 70434 NA NA NA
inH2O@
4 °C
27.681 830.42 2768.1 8304.2 13840 27681 41521 83042 NA
inH2O@
20 °C
27.730 831.89 2773.0 8318.9 13865 27730 41595 83189 NA
mmHg@
0 °C
51.715 1551.5 5171.5 15515 25858 51715 NA NA NA
inHg@
0 °C
2.0360 61.081 203.60 610.81 1018.0 2036.0 3054.0 6108.1 10180
Overload Protection
Fuseless overvoltage protection
Temperature Coefficient
0.005 % of range per °C for temperature ranges of between −10 °C to 18 °C and 28 °C to
55 °C.
Loop Supply
24 V dc nominal
Pressure Module Input
Determined by pressure module