71X Series
Calibration Manual
26
Table 34. Required Calibration Equipment (cont.)
Calibrator
Model
Equipment Minimum Specifications
Recommended
Model
714 DC Calibrator -10 to 75 mV
Accuracy: 0.006 % + 0.25 mV
Fluke 5520A
Reference
Multimeter
-10 to 75 mV
Accuracy: 0.006 % + 0.25 mV
Fluke 8508A
TC to Banana Test
Leads
Copper Mini-Jack to Copper Wire
Type J TC Test
Lead
5500A/LEADS
715 DC Calibrator 0 to 10 Volts
Accuracy: 0.005 % + 0.5 mV
Fluke 5520A
Reference
Multimeter
0 to 10 Volts
Accuracy: 0.005 % + 0.5 mV
Fluke 8508A
Two Test Leads
716 DC Calibrator 0 to 24 mA Fluke 5520A
Reference
Multimeter
0 to 24 mA
Accuracy: 0.006 % + 0.25 mA
Fluke 8508A
Two Test Leads 5500A/LEADS
700P Pressure
Module
Load Resistors 1 k & 2 k 10% 0.5 W
717 DC Calibrator 0 to 24 mA Fluke 5520A
Reference
Multimeter
0 to 24 mA
Accuracy: 0.006 % + 0.25 mA
Fluke 8508A
Dead Weight Tester -12 to 100 PSIG
Accuracy: 0.012 % of range
Two Test Leads 5500A/LEADS
700P Pressure
Module
Load Resistors 1 k & 2 k 10% 0.5W
718/718Ex DC Calibrator 0 to 24 mA Fluke 5520A
Reference
Multimeter
0 to 24 mA
Accuracy: 0.006 % + 0.25 mA
Fluke 8508A
Dead Weight Tester -12 to 100 PSIG
Accuracy: 0.012 % of range
Two Test Leads 5500A/LEADS
700P Pressure
Module
Load Resistors 1 k & 2 k 10% 0.5 W