Process Calibrators
Verification
37
Table 53. 715 DC Voltage Source Verification (0.000 to 10.000 V)
Fluke 715 Fluke 8508A
0.000 V -0.002 V to + 0.002 V
5.000 V 4.997 V to 5.003 V
10.000 V 9.996 V to 10.004 V
5. Press the V key on the 715 Calibrator. Display should change to:
OUTPUT 0.00 mV
6. Using the large [ key scroll button of the 715 Calibrator, press to step to the
voltages in Table 54, verifying the output on the Fluke 8508A.
Table 54. 715 DC Voltage Source Verification (0.00 to 100.00 V)
Fluke 715 Fluke 8508A
0.00 mV -0.02 mV to + 0.02 mV
50.00 mV 49.97 mV to 50.03 mV
100.00 mV 99.96 mV to 100.04 mV