42
Table 63. 715 DC Voltage Measure Verification (25.0000 to 0.0000 V)
Fluke 5520A Fluke 715
20.00000 V 19.996 V to 20.004 V
10.00000 V 9.997 V to 10.003 V
0.0000 V -0.002 V to +0.002 V
4. Press the V key on the 715 Calibrator. Display should indicate closely to:
INPUT 0.00 mV
5. Set the Fluke 5520A for the mV settings in Table 64, and verify display readings on
the 715 Calibrator.
Table 64. 715 DC Voltage Measure Verification (0.0000 mV to 200.0000 mV)
Fluke 5520A Fluke 715
0.0000 mV -0.02 mV to + 0.02 mV
50.0000 mV 49.98 mV to 50.02 mV
150.0000 mV 149.96 mV to 150.04 mV
6. Disconnect the test leads from 715 Calibrator and turn the green O key off.
The 715 verification test is complete.