71X Series
Calibration Manual
54
INPUT ----- Ω,
then change to:
INPUT 350
7. Set the Fluke 5520A to 350 Ω.
8. Press C. The display should read:
INPUT -----
then change to:
INPUT 500 .
9. Set the Fluke 5520A to 500 .
10. Press C. The display should read:
INPUT -----
then change to:
INPUT 3200
11. Set the Fluke 5520A to 3200 .
12. Press C. The display should read:
INPUT ----- .
then change to:
OUTPUT 17-50 .
13. Using four test leads, connect the four jacks of Fluke 8508A ( Sense 'Hi & Lo' and
Input 'Hi & Lo') to the two middle jacks of the 712 Calibrator (black to black; red to
red). This puts the Fluke 8508A in a 4-wire configuration.
14. Set the 3458A to the 1000 range.
15. Press ] and } until the 712 display matches the Fluke 8508A.
16. Press C. The display should read:
OUTPUT -----
then change to:
OUTPUT 350.00 0.
17. Press ] and } until the 712 display matches the Fluke 8508A.
18. Press C. The display should read:
OUTPUT -----
then change to:
OUTPUT 126.00
19. Put the Fluke 8508A in the 10 k range.
20. Press ] and } until the 712 display matches the Fluke 8508A.
21. Press C. The display should read:
OUTPUT -----
then change to:
OUTPUT 2535.0
22. Press ] and } until the 712 display matches the Fluke 8508A.
23. Press C. The display should read: