Introduction and Specifications
Specifications
1
1-5
Table 1-1. 8060A Specifications
Electrical
The following specifications are based on a one-year calibration cycle, an
operating temperature of 18 to 28°C (64 to 82°F) and a relative humidity
not exceeding 80%.
DC Voltage
Range Resolution
Accuracy
±(% of reading + no. of digits)
200 mV
2V
.01 mV
.1 mV
0.04% + 2
20V
200V
1000V
1 mV
10 mV
100 mV
0.05% + 2
Response Time............................. 1 second maximum, to rated
accuracy within selected range.
Input Impedance ........................... 10 MΩ nominal
Normal Mode Noise Rejection ...... >60 dB at 50 Hz or 60 Hz
Common Mode Noise Rejection ... >120 dB at dc, >90 dB at 50 Hz
and 60 Hz (1 kΩ imbalance)
Overload Protection ...................... 1000V dc or peak ac
continuous, except 20 seconds
maximum on 200 mV and 2V
ranges above 300V dc or rms.
DC Voltage, High Impedance Mode
All specifications are the same as for the dc voltage mode except
the following (only 200 mV and 2V ranges are available):
Range Resolutions
Accuracy
±(% of reading + no. of digits)
200 mV
2V
.01 mV
.1 mV
0.05% + 2
Input Impedance ........................... >1,000 MΩ, typically 10,000 MΩ
Overload Protection ...................... 300V dc or rms continuous, 20
seconds maximum 300V to
1000V dc or peak ac.