8060A
Instruction Manual
1-6
Table 1-1. 8060A Specifications (cont)
DC Voltage, dB Mode
Measurements are made in dBm referenced to 600 or relative dB. All
specifications are the same as for dc voltage except the following:
Dynamic Range............................. With full .01 dB resolution, 99.79
dB. Total specified dynamic
range is 136.22 dB (160 µV to
1000V).
Resolution and Accuracy .............. Depends on linear dc count
(count refers to the display in dc
volts independent of the decimal
points - see table below).
Accuracy
dBm Ref. 600
Range Tolerance
-74 to -56
(160 µV to 1.27 mV)
200 mV
±1 dB
-55.6 to -37.6
(1.28 mV to 10.23 mV)
200 mV
±.2 dB
-37.58 to -31.77
(10.24 mV to 19.99 mV)
200 mV
±.04 dB
-31.76 to -11.76
(20 mV to 199.99 mV)
200 mV
±.04 dB
-11.76 to 8.24
(.2V to 1.9999V)
2V
±.04 dB
8.24 to 28.24
(2.000V to 19.999V)
20V
±.04 dB
28.24 to 48.24
(20.00V to 199.99V)
200V
±.04 dB
48.24 to 62.22
(200.0V to 1000V)
1000V
±.04 dB
Linear Counts Resolution Accuracy
19.999 to 1024
1023 to 128
127 to 16
.01 dB
.1 dB
1 dB
±.04 dB
±.2 dB
±1 dB