Introduction and Specifications
Specifications
1
1-7
Table 1-1. 8060A Specifications (cont)
AC Voltage (True rms, AC-Coupled)
Ranges.................. 200 mV, 2V, 20V, 200V, 750V
Accuracy * ............. ±(% of reading + no. of digits). See table below:
Input
Voltage
Reso-
lution
Range
20 Hz -
45 Hz
45 Hz -
1 kHz
1 kHz -
10 kHz
10 kHz -
30 kHz
30 kHz -
50 kHz
50 kHz -
100 kHz
20.0 -
199.99 m
.01 mV 200 mV
0.2%
+ 12
0.2%
+ 20
0.5%
+ 40
1%
+ 100
.2000 -
1.9999V
.1 mV 2V
1% + 10
3% +200**
2.000 -
19.999V
1 mV 20V
0.5%
+12
0.5%
+20
1% +40 2% +100
20.00 -
199.99V
10 mV 200V
75.0 -
4.99.9V
100 mV 750V
Not Not Specified
500.0 -
750.0V
Specified
1%
+12
* Not specified for input signals <10% of range.
** For input voltage between 10% and 15% of range, add an additional 140 counts.
Input Impedance......... 10 MΩ shunted by <100 pF
Common Mode Noise
Rejection..................... >60 dB at 50 Hz and 60 Hz (1 kΩ
imbalance)
Crest Factor Range .... 1:1 to 3:1
Response Time........... Five seconds maximum to rated accuracy
within selected range, 12 seconds to rated
accuracy from an overload.
Overload Protection.... 750V rms or 1000V peak continuous
except 20 seconds maximum on the 200
mV range above 300V rms or 300V dc.
Input not to exceed a volt-hertz product of
10
7