8060A
Instruction Manual
1-8
Table 1-1. 8060A Specifications (cont)
AC Voltage, dB Mode (True rms, AC-Coupled)
Measurements are made in dBm referenced to 600 or relative dB. All
specifications are the same as for ac voltage except the following:
Dynamic Range... With full .01 dB resolution, 97.30 dBm. Total specified
dynamic range is 109.72 dBm (2.45 mV to 750.0V ac
rms).
Resolution ........... Depends on number of linear ac counts (count refers
to the actual number on the display independent of
the decimal point. See table below).
Linear Counts* Resolution
19.999 to 1024
1023 to 128
127 to 16
.01 dB
.1 dB
1 dB
*Not specified below 245 counts.
Accuracy................. See table below:
dBm Ref. 600 Range 20 Hz -
45 Hz**
45 Hz -
10 kHz**
10 kHz -
30 kHz**
30 kHz -
50 kHz**
50 kHz -
100 kHz**
-50.0 to -31.76
(2.45 mV to 20.00 mV)
200 mV Not Specified
-31.76 to -29.83
(20.00 mV to 25.00 mV)
200 mV 0.20 dB 0.50 dB 1.00 dB 2.70 dB
-29.83 to -11.76
(25.00 mV to 199.99 mV)
200 mV 0.10 dB 0.15 dB 0.30 dB 0.50 dB
-11.76 to 8.24
(.2000V to 1.9999V)
2V
0
.10 dB*0.10 dB* 0.15 dB* 0.30 dB* 0.50 dB*
8.24 to 28.24
(2.000V to 19.999V)
20V
28.24 to 48.24
(20.00V to 199.99V)
200V 0.15 dB 0.30 dB 0.30 dB 0.65 dB 1.83 dB
48.24 to 59.72 750V
20 Hz -
1 kHz**
1 kHz - 100 kHz**
(200.0V to 750.0V)
0.5 dB Not Specified
*Specification applies above 8000 linear counts.
**Not specified for input signals <10% of range.