EasyManua.ls Logo

Fluke 8060A - Page 21

Fluke 8060A
140 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Introduction and Specifications
Specifications
1
1-9
Table 1-1. 8060A Specifications (cont)
AC Voltage, dB Mode (cont.)
1
Hz
5
Hz
10
Hz
20
Hz
100
Hz
10
kHz
100 mV
Not Specified
Not Specified
+1
0
-1
-3
-2
+1
dB
(relative
to 200 Hz
reading)
0
-1
-3
-2
25 mV
50
kHz
100
kHz
200
kHz
300
kHz
400
kHz
Frequency
Typical Response in 200 mV Range
Frequency
Frequency Range
(Fully Autoranging)
Resolution
Accuracy
±
(% of reading + no. of digits)
200 Hz
2000 Hz
20 Hz
200 Hz
.01 Hz
.1 Hz
1 Hz
10 Hz
.05% + 1
Input Signal Sensitivity (based on sine wave V rms)
12 Hz to 20 kHz
20 kHz to 100 kHz
100 kHz to 200 kHz
20 mV or 10% of voltage range*
50 mV or 25% of voltage range*
150 mV or 75% of voltage range*
*Whichever value is greater.
Not Specified
5V
Voltage
(sine wave rms)
50 mV
20 mV
16 mV
150 mV
70020010020
Spec. Limit
Typical
Frequency (kHz)
Frequency Input Sensitivity (200 mV range)
"