Table 1-1. 8060A Specifications (cont)
Frequency (cont.)
AC Voltage Range
Maximum Useable AC
Voltage*
200 MV
2V
20V
200V
750V
±5V peak
±50V peak
±500V peak
±1000V peak
±1000V peak
*Signal not to exceed a volt-hertz product of 1 x 10
7
.
Input Characteristics ..... Ac-coupled, 10 M shunted by <100 pF
Overload Protection ...... 759V rms or 1000V peak continuous
except 20 seconds maximum on the 200
mV range above 300V rms or 300V dc.
Input not to exceed a volt-hertz product of
10
7
(for example, 200V at 50 kHz).
Extended Frequency
Selection................... Enabled by holding down Hz button at
power on.
Range....................... 12 Hz to 700 kHz, typically.
Resolution ................ 100 Hz above 200 kHz.
Accuracy................... ±(0.5% of reading + 2 digits)
Sensitivity
(sine wave V rms)..... Typically 100 mV at 200 kHz increasing to
4.5V at 700 kHz in the 200 mV range. Will
measure a TTL signal (50% duty cycle) to
420 kHz, typically.
Resistance
Ranges.......................... 200, 2 k, 20 k, 200 k, autoranging
M. The M range extends from .0001
m to 300 M in three autoranged
ranges. Upscale range changes are made
at 2 M and 20 M. Downscale range
changes are made at 19 M and 1.9 M.
Accuracy ....................... ±(% of reading + no. of digits). See table
below.