Introduction and Specifications
Specifications
1
1-11
Table 1-1. 8060A Specifications (cont)
Resistance (cont.)
Range
Reso-
lution
Accuracy
Full-
scale
Voltage
Max
Current
Open
Circuit
Voltage
200Ω 0.01Ω (0.07%+2+.02Ω) <1.1 mA <4.8V
2 kΩ 0.1Ω (0.07%+2)
<250 mV
<150 µA
20 kΩ 1Ω (0.07%+2) <15 µA <1.5V
200 kΩ 10Ω (0.07%+2) <1.5 µA
0-1.9999 MΩ 100Ω (0.15%+2)
MΩ
2-19.99 M
Ω 10 kΩ (0.2%+3) 2.5 µA
20-99.9 MΩ 100 kΩ (1%+3) <2.5V <2.5V
100-300 MΩ 1 MΩ (2%+3)
Autoranging
k
Ω
0.1Ω
to 1 kΩ
(0.2%+5) <1.0 mA
Response Time.............. Two seconds maximum to rated accuracy
for all ranges except MΩ. For MΩ, 8
seconds maximum.
Overload Protection....... 300V dc or rms ac for all ranges
Conductance
Range ............................ 2000 nS (equivalent to a resistance
range from 500 kΩ to 10,000 MΩ)
Resolution...................... 0.1 nS
Accuracy........................ ±(0.5% of reading + 20 digits)
Open Circuit Voltage...... <1.5V
Overload Protection....... 300V dc or rms ac