8060A
Instruction Manual
1-12
Table 1-1. 8060A Specifications (cont)
Continuity
Ranges.......................... All resistance ranges
Threshold ...................... Nominally 10% of range (for example,
20 in the 200 range) for 200, 2 k,
20 k, 200 k ranges. Nominally 20 k
in M range.
Display Indication.......... Horizontal bar across the top of the
display and/or 2.667 kHz tone. Indication
is present for a minimum of 200 ms.
Response Time............. 50 µs maximum (10 µs typical)
Overload Protection ...... 300V dc or rms ac
Diode Test
Range............................ 2V
Test Current .................. 1 mA (typical)
Accuracy ....................... ±(0.05% of reading + 2 digits)
(Specification applies for voltage
measurement)
Response Time............. 2 seconds maximum
Overload Protection ...... 300V dc or rms ac
DC Current
Range Resolution
Accuracy
±(% of reading + no. of
digits)
Burden
Voltage
200 µA
2 mA
.01 µA
.1 µA
0.2% + 2
.3V typical
.3V typical
20 mA
200 mA
2000 mA
1 µA
10 µA
100 µA
0.3% + 2
.3V typical
.3V typical
.9V typical
Overload Protection ......... 2A/250V fuse (operator replaceable) in
series with 3A/600V fuse (service
personnel replaceable).