Operation
Simulations
3
3-15
Standard BP
Preset # 1
Cuff: External
OPTIONS CUFF
ready ....
120/ 80 (93)
80 BPM 0.68 cc
fas07.eps
2. Press the OPTIONS soft key to scroll through the simulation choices.
Parameters for Standard BP simulations are shown in Table 3-2.
Table 3-2. Standard Blood Pressure Simulations
Simulation
Number
Blood Pressure
(mmHg) (MAP)
Heart Rate (bpm)
Pulse volume
(cc)
1 120/80 (93) 80 0.68
2 150/100 (116) 80 0.65
3 200/150 (166) 80 0.60
4 255/195 (215) 80 0.55
5 60/30 (40) 80 0.75
6 80/50 (60) 80 0.71
7 100/65 (76) 80 0.69
3. Press the CUFF soft key to select Internal Adult or External cuff.
4. Press Start on the NIBP unit.
5. Compare values on the NIBP unit with those on the Tester.
6. Press HOME to return to the Main menu.
Patient Conditions
The Patient Condition simulations are intended to provide some basic patient
variations.
To perform Patient Conditions simulations, do the following:
1. Access these simulations by pressing the 5 PATIENT CONDITIONS key.
The Patient Condition screen appears.