Operation
Auto Sequences
3
3-25
Select Auto Sequence: 1
fas42.eps
3. Press a numeric key (1-9), followed by the ENT key.
The auto sequence begins to execute, based on the options that were enabled
using the EDIT function.
Note
Tests and simulations not enabled in the executing auto sequence are
skipped.
Enabled auto sequences are executed in the following order:
Pressure Gauge Test
Pressure Leak Test
Pressure Relief Valve Test
Pressure Source Test
Data Sheet Printout
NIBP Simulations
Pressure Gauge
This test monitors system pressure, which must be generated external to the Tester.
Press the NEXT> key to advance to the next test. Only the pressure displayed
when the NEXT> key is pressed appears on the printout.
Leak Test
This test performs a system leak test at the Setpoint pressure, which was defined
when the auto sequence was edited.
1. Press the START soft key to initiate this test. The system then pressurizes.
Once a stable pressure reading near the Setpoint is reached, a 60-second timer
begins to count down. Once the timer has elapsed, a Leak Rate appears on the
display.

Table of Contents