3-24
Auto Sequence #1 Definition
Gauge Enabled
Leak Setpoint: 200 mmHg
Relief Pressure:
380 mmHg
Source Pressure: 200 mmHg
Cuff: External
Simulation 1: Preset #1 120/80
80BPM 0.7cc
Simulation 2: Preset #2 150/100
80BPM 0.7cc
Simulation 3: Preset #3 200/150
80BPM 0.7cc
Simulation 4: Preset #4 255/195
80BPM 0.7cc
Simulation 5: Preset #5 60/30
80BPM 0.7cc
Figure 3-7. Sample Auto Sequence Test Printout
Running Auto Sequences
Use the following steps to run an auto sequence.
1. Press the AUTO SEQUENCE soft key to access the Auto Sequence
menu.
Auto Sequence
EDIT PRINTRUN
fas36.eps
2. Press the RUN soft key. The Select Auto Sequence (1-9) screen appears.