BP Pump 2
Operators Manual
3-20
Wrist
The Wrist simulations are provided to test wrist cuff NIBP monitors.
To perform Wrist simulations, do the following:
1. Access these simulations by pressing the 9 WRIST key. The Wrist
Simulation screen appears.
Wrist Simulation
WristCuff # 1
Cuff: External
OPTIONS
ready ....
120/ 80 (93)
80 BPM 0.50 cc
fas13.eps
2. Press the OPTIONS soft key to scroll through the simulation choices. The
simulation is automatically set up to use the external cuff and cannot be
changed. Parameters for Wrist simulations are shown in Table 3-7.
Table 3-7. Wrist Simulations
Simulation
Number
Blood Pressure
(mmHg) (MAP)
Heart Rate (bpm)
Pulse volume
(cc)
1 120/80 (93) 80 0.50
2 160/100 (120) 80 0.50
3 80/55 (63) 80 0.50
7. Press Start on the NIBP unit.
8. Compare values on the NIBP unit with those on the Tester.
9. Press HOME to return to the Main menu.