BP Pump 2
Operators Manual
3-28
Fluke Biomedical
Tester Serial #:123456
Date:________ Time:_______
Serial #: ___________________
Control #: __________________
Mfg: ________________________
Model: ______________________
Location: ___________________
Technician: _________________
Work Order: _________________
Procedure ID: _______________
Gauge: 181 mmHg
Leak Test
Leak Rate: 5 mmHg/min
Start Pressure 203 mmHg
End Pressure: 198 mmHg
Relief Valve Test
Peak Pressure:
No Relief Detected
418 mmHg
Pressure Source
Actual:
Source:
_________
186 mmHg
Actual ___/___ ( ) ___BPM
Preset #1 120/80 80BPM 0.7cc
Actual ___/___ ( ) ___BPM
Preset #2 150/100 80BPM 0.7cc
Actual ___/___ ( ) ___BPM
Preset #3 200/150 80BPM 0.7cc
Actual ___/___ ( ) ___BPM
Preset #4 255/195 80BPM 0.7cc
Actual ___/___ ( ) ___BPM
Preset #5 60/30 80BPM 0.7cc
Figure 3-8. Sample Data Sheet Printout