Expert Series Thermal Imagers
Menus
15
Transmission <options>
Changes the transmission percentage of the infrared-
transparent window (IR window).
When you do infrared inspections through IR windows, not
all of the infrared energy emitted from the target is
transmitted through the optical material in the window. If
you know the transmission percentage of the window,
adjust the transmission percentage in the Imager or in
SmartView software to improve the accuracy of the
measurement.
Note
If Display is set to Display All, the
transmission correction shows as t = xxx% on
the display.
Spot Temp
Hot
Select to view and turn on/off either the hot or cold spot
indicator on the display.
The Spot Temperatures are floating HI and LO
temperature indicators that move on the display as the
temperature measurements of the image fluctuate.
Cold
Spot Markers
All Off Turns off fixed-temperature spot markers.
<options>
Select the number of fixed-temperature spot markers to
use to highlight a region before you take an image. See
Spot Markers.
Spot Box
On
Turns on/off a temperature measurement zone (box) that
centers on a target.
Off
Set Size
With Spot Box set to On, changes the size of the Spot
Box. See Spot Box.
Set Position
With Spot Box set to On, changes the position of the Spot
Box. See Spot Box.
Table 9. Measurement Menu (cont.)
Option Menu Option Description