Step Action
5 Remove the screw cap that covers the clamp ring screw (D) using the tool
i
llustrated.
6 Take the 8 mm HEX attachment head and fit it to the tool as illustrated be-
l
ow. Hold the pressure plate and while removing the four hex screws hold-
ing the pressure plate in place using the tool. Then remove the pressure
plate (E).
7 Wrap new PTFE tape around the threads of the four hex screws to prevent
t
hread seizure.
5 Preparation
5
.2 Install disposable flow kits
ÄKTA ready XL Operating Instructions 29281616 AE 101