Step Action
5 Carefully disconnect and lift off the air trap level sensor fixture, level sen-
s
ors, lever, and the strap from the disposable air trap. These items are re-
usable and must not be removed from the system
NOTICE
Do not remove the level sensor fixture, level sensors,
lever, and the strap from the system. These items are
r
eusable parts of the system.
6 Make sure to fit caps to all sensor connections on the system with the caps
p
rovided on the system.
7 Set the switch to Flowkit install to open the pinch valves.
8 Remove the flow kit sections starting from the outlet/column side by
opening the pinch valves locking pins one by one as the tubing is removed
from the valve. Close each valve after the tubing is removed.
9 Dispose of the flow kit sections in accordance with local regulations.
Shutdown procedure
T
he example below describes the shutdown procedure for UNICORN software.
For other control softwares, refer to the software documentation.
Step Action
1 In UNICORN, select F
ile →Exit UNICORN in any module, or select File
→Quit Program in UNICORN Manager, depending on the UNICORN ver-
sion used.
2 When the touchscreen monitor has switched off, turn off the Power
switch.
Note: For LOTO make sure to immobilize the power switch by inserting the pad-
l
ock into the power switch locking hole.
6 Operation
6
.3 Controlled shutdown and removal of flow kit
142
ÄKTA ready XL Operating Instructions 29281616 AE
*** TeDIS Valid on Date Printed Printed on: 15-09-2023 11:27:35 ***