Step Action
5 Carefully disconnect and lift off the air trap level sensor fixture, level sen-
s
ors, lever, and the strap from the disposable air trap. These items are re-
usable and must not be removed from the system
NOTICE
Do not remove the level sensor fixture, level sensors,
lever, and the strap from the system. These items are
r
eusable parts of the system.
6 Make sure to fit caps to all sensor connections on the system with the caps
p
rovided on the system.
7 Disconnect the flow kit sensors and then disconnect each flow kit section.
8 Set the switch to Flowkit install to open the pinch valves, and press Re-
set.
9 Remove each flow kit section by opening the pinch valves locking pins one
by one as the tubing is removed from the valve. Close each valve after the
tubing is removed.
10 Dispose of the flow kit sections in accordance with local regulations.
Alternative 3 - Drain and remove
the entire flow kit
To drain the flow kit, and then remove the entire flow kit follow the instructions be-
low.
Note: The inlets and pump sections are not entirely drained using this procedure,
t
herefore splashing is to be expected.
Step Action
1 Connect a drain tube to the C
olumn bottom outlet tubing and lead it to a
suitable waste container.
2 Close any valves to tanks leading to the system inlets.
3 Set the switch to Drain.
4 When no more waste is noticed, disconnect and remove customer tubing
from the Column bottom outlet.
6 Operation
6
.3 Controlled shutdown and removal of flow kit
ÄKTA ready XL Operating Instructions 29281616 AE 141