Step Action
1 Empty the air trap completely from liquid using the D
rain instruction for
the air trap (see Air trap valves, on page 58).
2 Use the instruction System →Calibrate →LSL166_air to calibrate the low
level sensor using air.
3 Use the instruction System →Calibrate →LSH167_air to calibrate the high
level sensor using air.
4 Fill the air trap until liquid comes out through the top valve by pumping in
liquid using the Fill instruction.
5 Turn off the pump.
6 Use the instruction System →Calibrate →LSL166_liquid to calibrate the
low level sensor using liquid.
7 Use the instruction System →Calibrate →LSH167_liquid to calibrate the
high level sensor using liquid.