Filter valves
T
he filter valve blocks are directly connected to the air trap valve blocks and to tubing
going to the air sensor prior to the column. If no customer filter is used, set the valves
to bypass. See the valve macros for the filter instruction options below.
Valve positions Open valves
Bypass XV-026
Inline
(default) XV-025 + XV-027
Column valves
T
he column valve set consist of five valves. See the valve macros for the column in-
struction options below.
Valve positions Open valves
Bypass XV-035
UpFlow XV-032
+ XV-033
DownFlow XV-031 + XV-034
Open Bottom XV-033 + XV-034
Open Top XV-031 + XV-032
3 System description
3
.4 System components
3.4.4 Valves
ÄKTA ready XL Operating Instructions 29281616 AE 59