9 System maintenance
68 GE Healthcare 86500-IMG rev 3
9.5 QUALITY CONTROL CHECK
If the performance of the unit has changed, please contact your local
distributor.
*) The serial number is printed on a label attached to the rear cover of the
unit.
Check for the operation of the C-Arm gear. Lubricate, if necessary.
Check the collimator and filter mechanism and movement.
Inspect the floor fixing points to ensure the secure and safe fixing of the
support base plate of the unit.
Check the stereo drive mechanism.
Check Tube Park alignment.
Update the tube efficiency software parameters.
Remove debris from the cassette loading mechanism and lubricate it.
Apply grease annually on Stereo drive mechanism if Delta 32 is used with
Diamond.
Emergency buttons Verify that all electronically controled
movements (vertical, c-arm rotation,
compression) are disabled when
emergency button is pressed down. All
motorized movements should stop.
Release the emergency stop button,
switch on the unit once again and then
repeat the test using the emergency stop
button on the opposite side of the unit.
Finally, verify that normal operation of
the unit is restored once the button is
back in normal position.
Distributor:
Contact person: Tel:
Diamond serial
number *):
Installation
date:
Last maintenance
performed:
Exposure counter
number:
Last maintenance
performed:
Exposure counter
number:
Once a year Procedure/Section in Diamond
Installation Instructions (86501-IMG)
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).