7 Menu options
56 GE Healthcare 86500-IMG rev 3
6 Press the Cancel key to return to normal operation.
NOTE!
If the test result differs considerably (max. 0.15-0.20, depending on
country-specific regulations) from the required O.D. value (depending on
country-specific regulations, e.g. 1.60), exit the test and check the film
processor.
7.2 OTHER MODES
7.2.1 Pre-Programmed/Specimen Modes
For the user convenience, Diamond software incorporates 4 pre-
programmed modes, identifiable by name up 8 digits long for easy recall.
Your service technician can set the following parameters for individual
mode:
–density
offset for initial kV in Auto kV mode
target time 1/2
name (max. 8 characters)
Auto kV change limit
AutoPoint coefficient
If the pre-programmed mode is renamed (max. 8 characters), the name of
the mode will be displayed in left upper corner of the Control Panel LCD-
screen.
A Specimen mode can be used to examine tissue that has been removed
from the patient. In Specimen mode, kV values range 15 to 39 kV. When
Specimen mode is selected, the LCD screen blinks and vertical movement
is disabled.
WARNING!
Use Specimen mode to examine only the removed tissue, not the patient
herself.
1 Select MODES from the menu screen:
The following screen appears:
2 Scroll up or down to the mode you want to select, then press Accept.