7 Menu options
86500-IMG rev 3 GE Healthcare 55
A No IQC adjustment necessary
1 If the IQC value is correct, press the Accept key to exit the QA Mode.
2 Press the Cancel key to return to normal operation.
B IQC adjustment necessary
If the test result shows a changed density reading, the current IQC value
needs to be adjusted to the newly measured density value.
1 Press the down and up keys until the newly measured density value
is displayed.
NOTE!
If the test result differs considerably (max. 0.15-0.20, depending on
country-specific regulations) from the required O.D. value (depending on
country-specific regulations, e.g. 1.60), exit the test and check the film
processor.
2 Press the middle button Accept to accept the value. The following
screen appears:
The screen display shows the new IQC coefficient with the remark Pass to
indicate that it is within the acceptable range.
3 Press the Cancel key to return to normal operation.
If the IQC value is not within the acceptable range of density values, the
following screen appears:
4 In this case, enter a new density value with the down and up keys
until the new value is displayed.
5 Press the Accept key to accept the value. The following screen
appears:
The screen display shows the new density value with the message Pass,
which indicates that it is within the acceptable range.