12 Diamond X-ray tube data sheets
86500-IMG rev 3 GE Healthcare 89
12.6 GENERAL DATA
12.7 B-112 HOUSING
Maximum Peak Voltage: 39 kV (Anode to Ground)
Nominal Focal Spot Size: Small 0.1 mm (Refer to General
Information Section for Focal Spot
Measurement Method)
Large 0.3 mm
Maximum Anode Load: 3-phase, 180Hz, 0.1 second
Small Focal Spot: 1320 W
Large Focal Spot: 2.6 kW
Anode Disc Construction: Molybdenum Alloy
Anode Disc Diameter: 77 mm
Anode Disc Target Angle: M113SP (LF/SF) 16/10
Anode Heat Storage: 300 kHU
Maximum Anode Cooling Rate: 60,000 HU/minute
Inherent Filtration: 0.0 mm Alum. Equiv. 0.76 mm
Beryllium
U/L Approved: Yes
Class: I
Type: B
Maximum Peak Voltage: 52 kV
Anode to Ground: 52 kV
Cathode to Ground: 0 kV
Housing Heat Storage Capacity: 370 KJ (500 kHU)
Housing Cooling Rate: 100 Watts (135HU/sec.)
Weight: 10.2 kg (22.5 lbs)
Length: 33 cm (13”)
Diameter: 13.8 cm (5.44”)
Focal Spot - Port Spacing: Small 47.0mm ± 1mm
Large 41.4 ± 1mm
High Voltage Cable, Special Design
(Non-Federal Standard):
55 kV Rating
Acoustic noise: 50dB Max. (A-weighted) at 30 cm
Leakage Technique Factor M113SP: 52 kV, 2mA
Leakage Technique Factor B112: 55 kV, 4mA
Minimum Ambient temperature for
storage & transportation:
-9°C (16°F)
Maximum Operation Temperature: +40°C (+104°F)
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).